Author:
Zhang Jun,Shao Jie,Zheng Li,Shen Ye,Zhao Xia
Abstract
Abstract
Background
Intraocular lens (IOL) calculation using traditional formulas for post-corneal refractive surgery eyes can yield inaccurate results. This study aimed to compare the clinical accuracy of the newly developed Zhang & Zheng (ZZ) formula with previously reported IOL formulas.
Study design
Retrospective study.
Methods
Post-corneal refractive surgery eyes were assessed for IOL power using the ZZ, Haigis-L, Shammas, Barrett True-K (no history), and ray tracing (C.S.O Sirius) IOL formulas, and their accuracy was compared. No pre-refractive surgery information was used in the calculations.
Results
This study included 38 eyes in 26 patients. ZZ IOL yielded a lower arithmetic IOL prediction error (PE) compared with ray tracing (P = 0.04), whereas the other formulas had values like that of ZZ IOL (P > 0.05). The arithmetic IOL PE for the ZZ IOL formula was not significantly different from zero (P = 0.96). ZZ IOL yielded a lower absolute IOL PE compared with Shammas (P < 0.01), Haigis-L (P = 0.02), Barrett true K (P = 0.03), and ray tracing (P < 0.01). The variance of the mean arithmetic IOL PE for ZZ IOL was significantly smaller than those of Shammas (P < 0.01), Haigis-L (P = 0.03), Barrett True K (P = 0.02), and ray tracing (P < 0.01). The percentages of eyes within ± 0.5 D of the target refraction with the ZZ IOL, Shammas, Haigis-L, Barrett True-K, and ray-tracing formulas were 86.8 %, 45.5 %, 66.7 %, 73.7 %, and 50.0 %, respectively (P < 0.05 for Shammas and ray tracing vs. ZZ IOL).
Conclusions
The ZZ IOL formula might offer superior outcomes for IOL power calculation for post-corneal refractive surgery eyes without prior refractive data.
Publisher
Springer Science and Business Media LLC
Subject
Ophthalmology,General Medicine
Cited by
4 articles.