Author:
Alqahtani Ali Robaian,Desai Shrikar R.,Patel Jignesh R.,Alqhtani Nasser Raqe,Alqahtani Abdullah Saad,Heboyan Artak,Fernandes Gustavo V. O.,Mustafa Mohammed,Karobari Mohamed Isaqali
Abstract
Abstract
Background
Dental implants emerge as a dependable and efficacious alternative for patients experiencing partial or complete tooth loss. The stability of these implants is influenced by surface topography and macro-level design. In cases where the height of the maxillary posterior region is diminished, employing short implants can prove advantageous. With the aim of examining the distribution of von Mises stress, strain, and micromovement in D4 bone quality surrounding platform-switched short implants, measuring 6 mm in length and featuring diameters ranging from 4 to 6 mm, as well as different thread designs, an in-depth finite element analysis was conducted under immediate loading conditions.
Methodology
A 3D finite element model was constructed to simulate maxillary molar crowns, incorporating an implant with a length of 6 mm and varying diameters and thread designs. The diameters utilized were 4/3.6 mm, 5/4 mm, and 6/4.8 mm, while the thread designs included buttress, square, and triangle patterns. Each model underwent analysis with a 100 N force applied in two directions: vertical and oblique, relative to the long axis of the implant. Stress, strain, and micromovement in the peri-implant region were recorded, employing the Ansys Workbench R v.18.1 software for modelling and analysis.
Results
When comparing all three diameters, the wide diameter (6 mm threads) exhibited the lowest values of peri-implant von Mises stresses (3.3 MPa and 35.1 MPa), strains (194 Ɛ and 484 Ɛ), and micromovements (0.7 μm and 1.3 Ɛ) subjected to axial and non-axial loading of a 100 N force. Notably, square microthreads yielded the most favorable stress parameters among the different thread shapes, manifesting the minimum values of stress, strains, and micromovements in their vicinity.
Conclusion
For the treatment of atrophic ridges or in scenarios necessitating extensive surgical preparation of the implant site, a combination of short implants, wide diameters, and platform switching can be employed. In situations with reduced bone height and the requirement for an implant-supported prosthesis to replace a missing permanent maxillary molar, the utilization of wide-diameter platform-switched short implants measuring 6 mm in length, featuring a square thread design, should be taken into consideration.
Publisher
Springer Science and Business Media LLC
Reference66 articles.
1. Sahin S, Cehreli MC, Yalcın E. The influence of functional forces on the biomechanics of implant-supported prostheses—a review. J Dent. 2002;30(7–8):271–82.
2. Belser U, Buser D, Bernard JP. Implants in load carrying part of dentition. In: Lindhe J, Lang NP, Salvi G, editors. Textbook of Clinical Periodontology and Implant Dentistry. 5th ed. Copenhagen: Blackwell Munksgaard; 2008. p. 954–61.
3. Araujo MG, Lindhe J. Peri-implant health. J Periodontol. 2018;89(Suppl 1):249–S256. https://doi.org/10.1002/JPER.16-0424.
4. Lin C-L, Kuo Y-C, Lin T-S. Effects of dental implant length and bone quality on biomechanical responses in bone around implants: a 3-D non-linear finite element analysis. Biomed Eng Appl Basis Commum. 2005;17:44–9.
5. Brunski JB. In vivo bone response to biomechanical loading at the bone/dental implant interface. Adv Dent Res. 1999;13:99–119.
Cited by
8 articles.
订阅此论文施引文献