Abstract
Abstract
Background
To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).
Methods
Eighty-two healthy right eyes were enrolled in this prospective observational study. Measurements were repeated for three times using the ZW-30 and IOLMaster 700 in a random order. Obtained parameters included axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism magnitude (AST), vector J0, vector J45, and corneal diameter (CD). The within-subject standard deviation (Sw), test–retest (TRT) variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were adopted to assess the intraobserver repeatability and interobserver reproducibility. The double-angle plot was also used to display the distribution of AST. To estimate agreement, Bland–Altman plots were used.
Results
For the intraobserver repeatability and interobserver reproducibility, the Sw, TRT and CoV for all parameters were low. Meanwhile, the ICC values were all close to 1.000, except for the J45 (ICC = 0.887 for the intraobserver repeatability). The double-angle plot showed that the distribution of AST measured by these two devices was similar. For agreement, the Bland–Altman plots showed narrow 95% limits of agreements (LoAs) for AL, CCT, AQD, ACD, LT, Km AST, J0, J45, and CD (− 0.02 mm to 0.02 mm, − 7.49 μm to 8.08 μm, − 0.07 mm to 0.04 mm, − 0.07 mm to 0.04 mm, − 0.07 mm to 0.08 mm, − 0.16 D to 0.30 D, − 0.30 D to 0.29 D, − 0.16 D to 0.16 D, − 0.23 D to 0.13 D, and − 0.39 mm to 0.10 mm, respectively).
Conclusions
The newly dynamic real-time visualization biometer exhibited excellent intraobserver repeatability and interobserver reproducibility. The two devices both based on the SS-OCT principle had similar ocular parameters measurement values and can be interchanged in clinical practice.
Funder
Project of National Natural Science Foundation of China
Shanghai Science and Technology
EYE & ENT Hospital of Fudan University High-level Talents Program
Clinical Research Plan of SHDC
Project of Shanghai Xuhui District Science and Technology
Program for Professor of Special Appointment at Shanghai Institutions of Higher Learning
Fondazione Roma and the Italian Ministry of Health
Publisher
Springer Science and Business Media LLC