Ab externo implantation of the MicroShunt, a poly (styrene-block-isobutylene-block-styrene) surgical device for the treatment of primary open-angle glaucoma: a review

Author:

Sadruddin OmarORCID,Pinchuk Leonard,Angeles Raymund,Palmberg Paul

Abstract

AbstractTrabeculectomy remains the ‘gold standard’ intraocular pressure (IOP)-lowering procedure for moderate-to-severe glaucoma; however, this approach is associated with the need for substantial post-operative management. Micro-invasive glaucoma surgery (MIGS) procedures aim to reduce the need for intra- and post-operative management and provide a less invasive means of lowering IOP. Generally, MIGS procedures are associated with only modest reductions in IOP and are targeted at patients with mild-to-moderate glaucoma, highlighting an unmet need for a less invasive treatment of advanced and refractory glaucoma. The PRESERFLO® MicroShunt (formerly known as InnFocus MicroShunt) is an 8.5 mm-long (outer diameter 350 μm; internal lumen diameter 70 μm) glaucoma drainage device made from a highly biocompatible, bioinert material called poly (styrene-block-isobutylene-block-styrene), or SIBS. The lumen size is sufficiently small that at normal aqueous flow hypotony is avoided, but large enough to avoid being blocked by sloughed cells or pigment. The MicroShunt achieves the desired pressure range in the eye by draining aqueous humor from the anterior chamber to a bleb formed under the conjunctiva and Tenon’s capsule. The device is implanted ab externo with intraoperative Mitomycin C via a minimally invasive (relative to incisional surgery) surgical procedure, enabling precise control of placement without the need for gonioscopy, suture tension control, or suture lysis. The implantation procedure can be performed in combination with cataract surgery or as a standalone procedure. The MicroShunt received Conformité Européenne (CE) marking in 2012 and is intended for the reduction of IOP in eyes of patients with primary open-angle glaucoma in which IOP remains uncontrolled while on maximum tolerated medical therapy and/or in which glaucoma progression warrants surgery. Three clinical studies assessing the long-term safety and efficacy of the MicroShunt have been completed; a Phase 3 multicenter, randomized clinical study comparing the MicroShunt to primary trabeculectomy is underway. In preliminary studies, the MicroShunt effectively reduced IOP and use of glaucoma medications up to 3 years after implantation, with an acceptable safety profile. This article summarizes current literature on the unique properties of the MicroShunt, the preliminary efficacy and safety findings, and discusses its potential use as an alternative to trabeculectomy for glaucoma surgery.

Publisher

Springer Science and Business Media LLC

Reference30 articles.

1. Gedde SJ, Feuer WJ, Shi W, Lim KS, Barton K, Goyal S, et al. Treatment outcomes in the primary tube versus trabeculectomy study after 1 year of follow-up. Ophthalmology. 2018;125(5):650–63.

2. Jones JP, Fong DS, Fang EN, Mesirov CA, Patel V. Characterization of glaucoma medication adherence in Kaiser Permanente Southern California. J Glaucoma. 2016;25(1):22–6.

3. Gedde SJ, Schiffman JC, Feuer WJ, Herndon LW, Brandt JD, Budenz DL. Treatment outcomes in the Tube Versus Trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol. 2012;153(5):789–803.e2.

4. Saheb H, Ahmed II. Micro-invasive glaucoma surgery: current perspectives and future directions. Curr Opin Ophthalmol. 2012;23(2):96–104.

5. Richter GM, Coleman AL. Minimally invasive glaucoma surgery: current status and future prospects. Clin Ophthalmol. 2016;10:189–206.

Cited by 57 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献