Author:
Puthettu Mira,Vandenberghe Stijn,Demertzis Stefanos
Abstract
Abstract
Background
During cardiac surgery, micro-air emboli regularly enter the blood stream and can cause cognitive impairment or stroke. It is not clearly understood whether the most threatening air emboli are generated by the heart-lung machine (HLM) or by the blood-air contact when opening the heart. We performed an in vitro study to assess, for the two sources, air emboli distribution in the arterial tree, especially in the brain region, during cardiac surgery with different cannulation sites.
Methods
A model of the arterial tree was 3D printed and included in a hydraulic circuit, divided such that flow going to the brain was separated from the rest of the circuit. Air micro-emboli were injected either in the HLM (“ECC Bubbles”) or in the mock left ventricle (“Heart Bubbles”) to simulate the two sources. Emboli distribution was measured with an ultrasonic bubble counter. Five repetitions were performed for each combination of injection site and cannulation site, where air bubble counts and volumes were recorded. Air bubbles were separated in three categories based on size.
Results
For both injection sites, it was possible to identify statistically significant differences between cannulation sites. For ECC Bubbles, axillary cannulation led to a higher amount of air bubbles in the brain with medium-sized bubbles. For Heart Bubbles, aortic cannulation showed a significantly bigger embolic load in the brain with large bubbles.
Conclusions
These preliminary in vitro findings showed that air embolic load in the brain may be dependent on the cannulation site, which deserves further in vivo exploration.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,General Medicine,Surgery,Pulmonary and Respiratory Medicine
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献