Gotfredsen, Anders, Lene Bæksgaard, and Jannik Hilsted.Body composition analysis by DEXA by using dynamically changing samarium filtration. J. Appl. Physiol.82(4): 1200–1209, 1997.—Dual-energy X-ray absorptiometry (DEXA) has a high accuracy for body composition analysis but is influenced by beam hardening and other error sources in the extremes of measurement. To compensate for beam hardening, the Norland XR-36 introduces a dynamically changing samarium filtration system, which depends on the current-absorber thickness. With this system we found a good agreement ( r = 0.99) between reference and measured amounts of tissue or fat percentages in a plastic phantom and in smaller (∼0.5–4 kg) and larger (∼5–20 kg) piles of tissue (ox muscle and lard). Scans of six healthy volunteers covered with combinations of beef and lard (∼5–15 kg) showed a good agreement ( r = 0.99) between reference and DEXA values of added soft tissue mass and fat percentage. We conclude that the DEXA method (and, in particular, the Norland XR-36 using dynamic filtration) has a high accuracy for body composition analysis. It has a potential for gaining status as a reference method in the future and may presently be used as a supplement to the traditional methods for body composition analysis.