A Preliminary Proposal for Quality Control Assessment and Harmonization of Leukocytes Morphology-Structural Parameters (Cell Population Data Parameters)

Author:

Seghezzi Michela,Buoro Sabrina,Previtali Giulia,Moioli Valentina,Manenti Barbara,Simon-Lopez Ramon,Ottomano Cosimo,Lippi Giuseppe

Abstract

SummaryBackground:The cell population data (CPD) measured by Sysmex XN-9000 can be used for screening many hematological and non-hematological disorders. Since little information is available on harmonization of CPD among different instrumentation and clinical laboratories, this study aimed at assessing the current degree of CPD harmonization between separate Sysmex XN modules allocated to the same laboratory.Methods:A total number of 78291 data were used for verification of within-run imprecision, analyzers harmonization, reference ranges and assessment of blood sample stability of CPD parameters, including results of daily quality control testing and those generated in samples collected from blood donors and healthy volunteers.Results:Within-run imprecision of CPD parameters ranged between 0.4 and 14.1%. Good agreement was found among five different XN-modules, especially when values were adjusted after calculation of instrument-specific alignment factors. The bias of all parameters remained always lower than the reference change values in samples stored for up to 8 hours, regardless of storage temperature.Conclusions:The imprecision of CPD parameters was acceptable, except for those reflecting the dispersion of cellular clusters. Due to the lack of reference control materials, we showed that the use of data generated on a large number of normal routine samples (i.e., a Moving Average population) may be a reliable approach for testing analyzers harmonization. Nevertheless, availability of both calibration and quality control materials for these parameters is highly advisable in the future. We finally showed that whole blood samples may be stable for up to 2–4 hours for most CPD parameters.

Publisher

Centre for Evaluation in Education and Science (CEON/CEES)

Reference40 articles.

1. PH Analytical goal setting for monitoring patients when two analytical methods are used;Petersen;Clin Chem,1992

2. Additional Tables for Steel Fligner Distribution - Free Multiple Comparisons of Three Treatments in and;Spurier;Communications Statistics Simulation Computation,2007

3. Novel Parameters in Counters;Lecompte;Blood Cell Clin Lab Med

4. Laboratory Standards Institute Evaluation of Precision of Quantitative Measurement Procedures approved guideline third edition document Laboratory Standards Institute;CLSI;Clinical Clinical,2014

5. The utility of the analyzer s parameters for detecting neutrophil dysplasia in myelodysplastic syndromes;Furundarena;Int J Lab Hematol,2010

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3