Probing deep lung regions using a new 6-electrode tetrapolar impedance method

Author:

Mobarak Mahjabin1,Kadir Muhammad Abdul2,Siddique-e Rabbani K2

Affiliation:

1. Bangladesh University of Professionals , Dhaka , Bangladesh

2. Department of Biomedical Physics and Technology, University of Dhaka , Dhaka , Bangladesh

Abstract

Abstract Probing deep regions of the lung using electrical impedance is very important considering the need for a low cost and simple technique, particularly for the low and medium income countries. Because of complexity and cost, Electrical Impedance Tomography is not suitable for this envisaged application. The simple Tetrapolar Impedance Measurement (TPIM) technique employing four electrodes is the age old technique for bioelectrical measurements. However, it has its limitations in respect of organ localisation and depth sensitivity using skin surface electrodes. Recently, a new 6-electrode TPIM with two current electrodes but two pairs of appropriately connected potential electrodes positioned on the front and back of the thorax, proposed by one of the authors, came with a promise. However, this work gave a qualitative proposal based on concepts of physics and lacked a quantitative evaluation. In order to evaluate the method quantitatively, the present work employed finite element method based COMSOL Multiphysics software and carried out simulation studies using this new 6-electrode TPIM and compared the results with those from 4-electrode TPIM, with electrodes applied either on the front or at the back of the thorax for the latter. Initially, it carried out a sensitivity distribution study using a simple rectangular volume conductor which showed that the 6-electrode TPIM gives better depth sensitivity throughout the lung region. Next it used a near life like thorax model developed by another of the authors earlier. Using this model, extensive studies were carried out to quantify the overall sensitivity over a target lung region, the contribution of the target lung to the total measured impedance, and several other parameters. Through these studies, the 6-electrode TPIM was established on a stronger footing for probing deep regions of the lungs.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering,Biophysics

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3