Optimization of Whole-Body CT Examinations of Polytrauma Patients in Comparison with the Current Diagnostic Reference Levels

Author:

Schäfer Stefan B.1,Rudolph Claudia2,Kolodziej Martin3,Mauermann Frank4,Roller Fritz Christian1,Krombach Gabriele Anja1

Affiliation:

1. Department of Diagnostic and Interventional Radiology, University-Hospital Giessen, Justus-Liebig-University Giessen, Germany

2. Department of Pediatric Radiology, University-Hospital Giessen, Justus-Liebig-University Giessen, Germany

3. Product Management, Infinitt Europe GmbH, Frankfurt, Germany

4. Clinical Application Specialist, Siemens Healthcare GmbH, Erlangen, Germany

Abstract

Purpose Evaluation of the dose values of a polytrauma whole-body CT examination used in clinical practice with regard to the 2016 updated diagnostic reference levels and reduction of the mean exposure levels using simple optimization steps. Materials and Methods In each case, 100 exposure values before and after dose optimization were compared with the old and new diagnostic reference levels. The grayscale values and the signal-to-noise ratio (SNR) were determined for the lung, the aortic arch and the liver. A visual assessment of the image quality was performed by two radiologists on the basis of a Likert scale (0 – non-diagnostic, 1 – poor visualization, 2 – moderate visualization, 3 – good visualization, 4 – excellent visualization) for CT examinations both before and after optimization. Results The acquired exposure values after dose optimization were below the old and new diagnostic reference levels (1319.98 ± 463.16 mGy · cm) while the mean value of the exposure values before optimization (1774.96 ± 608.78 mGy · cm) exceeded the current diagnostic reference levels. The measured grayscale values (HU) were (before versus after optimization): lung – 833 HU vs. – 827 HU (p = 0.43), aortic arch 341 HU vs. 343 HU (p = 0.70) and liver 68 HU vs. 67 HU (p = 0.35). After dose optimization the SNR in the lung was minimally higher, while it was minimally lower in the two other regions than before the optimization. Visual assessment of the image quality showed almost identical values with 3.85 evaluation points before and 3.82 evaluation points after dose optimization (p = 0.57). Conclusion Due to the updating of the diagnostic reference levels, an analysis of the own exposure values is necessary in order to be able to detect high values promptly and to initiate appropriate measures for dose reduction. Appropriate adaptation of the examination parameters with consideration of the necessary image quality allows a significant reduction of the radiation exposure in most cases, also on CT devices of older generations. Key Points: Citation Format

Publisher

Georg Thieme Verlag KG

Subject

Radiology Nuclear Medicine and imaging

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3