Simultaneous Frame-assisted Stereotactic Placement of Subdural Grid Electrodes and Intracerebral Depth Electrodes

Author:

Reinacher Peter C.12,Altenmüller Dirk-Matthias23,Krüger Marie T.24,Schulze-Bonhage Andreas23,Urbach Horst25,Trippel Michael12,Coenen Volker A.12,Delev Daniel24

Affiliation:

1. Department of Stereotactic and Functional Neurosurgery, Freiburg University Medical Center, Freiburg, Germany

2. Medical Faculty, Freiburg University, Freiburg, Germany

3. Department of Epileptology, Freiburg University Medical Center, Freiburg, Germany

4. Department of Neurosurgery, Freiburg University Medical Center, Freiburg, Germany

5. Department of Neuroradiology, Freiburg University Medical Center, Freiburg, Germany

Abstract

Background and Study Aims In complex cases of drug-resistant focal epilepsy, the precise localization of the epileptogenic zone requires simultaneous implantation of depth and subdural grid electrodes. This study describes a new simple frame-assisted method that facilitates the simultaneous placement of both types of intracranial electrodes. Material and Methods Ten consecutive patients were evaluated and divided into two groups. Group A included patients with simultaneous frame-assisted placement of depth and subdural grid electrodes. In group B, depth electrodes were implanted stereotactically; grid electrodes were implanted in a separate surgery. Results The placement of the subdural grid was accurate as individually designed by the epileptologists in all five patients from group A. In group B, one patient showed a slight and another one a significant deviation of the subdural grid position postoperatively. The mean surgical time in group A was shorter (280±62 minutes) compared with the mean duration of the surgical procedures in group B (336±51 minutes). Conclusion The frame-assisted placement of subdural grid electrodes facilitates the surgical procedure for invasive video-electroencephalography monitoring in complex cases of drug-resistant focal epilepsy in which a combination of depth electrodes and subdural grid electrodes is needed, by reducing the surgical time and guaranteeing highly accurate electrode localizations.

Publisher

Georg Thieme Verlag KG

Subject

Neurology (clinical),Surgery

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A multiple closed‐loops robotic calibration for accurate surgical puncture;The International Journal of Medical Robotics and Computer Assisted Surgery;2021-03-26

2. Single-Institutional Experience of Chronic Intracranial Electroencephalography Based on the Combined Usage of Subdural and Depth Electrodes;Brain Sciences;2021-02-28

3. Presurgical Evaluation;Surgical Treatment of Epilepsies;2020