Affiliation:
1. Cardiovascular Surgery Clinic, Memphis, Tennessee, United States
Abstract
Background The Cor-Knot device is a titanium fastener which is widely used in cardiac valve procedures. Adverse events associated with use of the device have been rare.
Case Description A patient underwent robotic mitral valve repair using the Cor-Knot fastener. Five years later, he suffered a cerebral vascular accident secondary to embolization of a metallic foreign body. He had had no other medical procedures which could account for a source of metallic embolization.
Conclusion Long-term surveillance after use of the Cor-Knot fastener is limited. Metallic embolization may be possible with this device and warrants awareness within the medical community.