Abstract

Abstract Purpose Due to the potential risk of COVID-19 transmission during endonasal surgery, we studied methods to reduce droplet and aerosol generation during these procedures. Methods Droplet spread was assessed using ultraviolet light and a camera that detected fluorescence in the operative field and surgeon's personal protective equipment. Density of aerosols sized <10 µm was measured using a photometric particle counter. We designed a face-mounted negative-pressure mask placed on the patients' face during endoscopic endonasal surgery. Sixteen patients were recruited between October 2020 and March 2021 and randomly assigned to the mask and no-mask groups. We compared droplet spread and aerosols generated in both groups, with copious irrigation and continuous suction during drilling forming the mainstay of surgical technique in all cases. Results Droplet contamination due to direct spillage of fluorescein from the syringe was noted in two patients. Aerosol density rose during sphenoid drilling in both groups, with no significant difference when continuous suction and irrigation were employed (1.27 times vs. 1.07 times the baseline, p = 0.248). Aerosol density rose significantly when suction and irrigation were interrupted in the no-mask group (44.9 times vs. 1.2 times, p = 0.028), which was not seen when the mask was used. Conclusion Aerosol generation increases during drilling in endonasal procedures and is a concern during this pandemic. The use of a rigid suction close to the drill along with copious irrigation is effective in reducing aerosol spread. The use of a negative pressure mask provides additional safety when inadvertent blockage of suction and inadequate irrigation occur.

Publisher

Georg Thieme Verlag KG

Subject

Neurology (clinical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3