Performance of a Maternal Abdominal Surface Electrode System for Fetal Heart Rate and Uterine Contraction Monitoring from 34 to 37 Weeks

Author:

Gonzalez Maritza1,Hill Meghan1,Cohen Wayne R.1

Affiliation:

1. Department of Obstetrics and Gynecology, University of Arizona College of Medicine, Tucson, Arizona

Abstract

Objective The objective of this study was to compare performance of a maternal surface electrode patch with ultrasound- and tocodynamometer-based monitoring to detect fetal heart rate and uterine contractility in late preterm labors. Study Design Thirty women between 340/7 and 366/7 weeks' gestation were monitored simultaneously with a Doppler/tocodynamometer system and a wireless fetal–maternal abdominal surface electrode system. Fetal and maternal heart rate and uterine contraction data from both systems were compared. Reliability was measured by the success rate and percent agreement. Deming regression and Bland–Altman analysis estimated the concordance between the systems. Uterine contractions were assessed by visual interpretation of monitor tracings. Results The success rate for the surface electrode system was 89.5% (95% confidence interval [CI], 85.7–93.3), and for ultrasound it was 88.4% (95% CI, 84.9–91.9; p = 0.73), with a percent agreement of 88.1% (95% CI, 84.2–92.8). Results were uninfluenced by the patients' body mass. The mean Deming slope was 1 and the y-intercept was −3.0 beats per minute (bpm). Bland–Altman plots also showed a close relationship between the methods, with limits of agreement less than 10 bpm. The percent agreement for maternal heart rate was 98.2% (95% CI, 97.4–98.8), and for uterine contraction detection it was 89.5% (95% CI, 85.5–93.4). Conclusion Fetal heart rate and uterine contraction monitoring at 340/7 to 366/7 weeks using abdominal surface electrodes was not inferior to Doppler ultrasound/tocodynamometry for fetal–maternal assessment. Registration clinicaltrials.gov/February 20, 2017/identifier NCT03057275. Key Points

Funder

Monica Healthcare, Ltd., Nottingham, UK

Publisher

Georg Thieme Verlag KG

Subject

Obstetrics and Gynaecology,Pediatrics, Perinatology, and Child Health

Reference28 articles.

1. Global burden of prematurity;M S Harrison;Semin Fetal Neonatal Med,2016

2. Use of continuous electronic fetal monitoring in a preterm fetus: clinical dilemmas and recommendations for practice;K Afors;J Pregnancy,2011

3. ACOG Practice Bulletin No. 106: Intrapartum fetal heart rate monitoring: nomenclature, interpretation, and general management principles;ACOG Practice Bulletin No;Obstet Gynecol,2009

4. The use of intrapartum electronic fetal heart rate monitoring: a national survey;D Devane;Ir Med J,2007

5. Obesity's impact on intrapartum electronic fetal monitoring;B Brocato;J Matern Fetal Neonatal Med,2019

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3