What Is the Cost of Off-Axis Insertion of Locking Screws? A Biomechanical Comparison of a 3.5 mm Fixed-Angle and 3.5 mm Variable-Angle Stainless Steel Locking Plate Systems

Author:

Kaczmarek Jakub1,Bartkowiak Tomasz2ORCID,Paczos Piotr2,Zawadzki Paweł2,Łączna Daria2,Gapiński Bartosz2

Affiliation:

1. Tierklinik Elversberg, Elversberg, Germany

2. Poznan University of Technology, Faculty of Mechanical Engineering, Poznan, Poland

Abstract

Abstract Objective The aim of this study was to evaluate the effect of screw insertion angle and insertion torque on the mechanical properties of a 3.5 fixed-angle locking plate locking compression plate (LCP) and 3.5 variable-angle locking plate polyaxial locking system (PLS). Methods In the LCP group, screws were placed abaxially at 0, 5 and 10 degrees. In the PLS group, screws were placed at 0, 5, 10, 15 and 20 degrees abaxially. The insertion torque was set to 1.5 and 2.5 Nm in the LCP and PLS groups respectively. A load was applied parallel to the screw axis, and the screw push-out force was measured until the locking mechanism was loosened. Results The 3.5 LCP showed higher push-out strength than the 3.5 PLS when the screws were placed at 0 degree regardless of the insertion torque. The off-axis insertion of 3.5 LCP locking screws resulted in a significant decrease in push-out strength (p < 0.05). A higher insertion torque value significantly increased the screw holding strength for the 3.5 LCP (p < 0.05). The 3.5 PLS system had a significantly higher push-out force when the screws are at 0 degree than at 5, 10 and 15 degrees, and 20 degrees (p < 0.05) at any given insertion torque. An increase in the insertion torque did not have a significant effect on the push-out strength of the 3.5 PLS locking system. Conclusion The 3.5 PLS is more sensitive to the screw insertion angle than to the insertion torque, whereas the 3.5 LCP is affected by both factors. Placing 3.5 LCP locking screws off-axis significantly reduces the screw holding strength; therefore, this approach has to be avoided. The findings of our research indicate that a 1.5 Nm torque can be used for a 3.5 PLS.

Funder

Ministry of Education and Science in Poland

Publisher

Georg Thieme Verlag KG

Subject

General Veterinary,Animal Science and Zoology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3