Affiliation:
1. Institute of Medical Microbiology and Hygiene, Saarland University, Homburg/Saar, Germany
2. Department of Thoracic and Cardiovascular Surgery, Saarland University Medical Center, Homburg/Saar, Germany
Abstract
Abstract
Mycobacterium chimaera is a nontuberculous mycobacterium that was identified as causative agent in a global outbreak of severe infections following open-chest cardiothoracic surgery. Heater-cooler units (HCUs), which were contaminated during the manufacturing process, were elucidated as the origin of this outbreak. Regular surveillance of water-containing HCUs used for cardiac surgery was recommended as one preventive measure. We present data on the occurrence of M. chimaera and other mycobacterial pathogens in different HCUs from one surgical center in Germany over a 42-month period. Water samples and swabs from seven different HCUs were taken between 2015 and 2018, and mycobacteria were detected in 50.6% (78/154) of water samples and 21.1% (4/19) of swabs. M. chimaera accounted for the majority of detected pathogens (77/83 isolates in water samples), but other species such as Mycobacterium gordonae were also found. Despite strict adherence to an intensified, regular disinfection procedure, the majority of HCUs remained positive for mycobacteria until the end of the study. In conclusion, additional measures are needed to reduce the risk of intraoperative transmission of M. chimaera, and our observations underscore the inherent infections risks associated with water-containing medical devices.
Subject
Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献