Effect of Blood Collection Tubes on Total Triiodothyronine and Other Laboratory Assays

Author:

Bowen Raffick AR1,Chan Yung1,Cohen Joshua1,Rehak Nadja N1,Hortin Glen L1,Csako Gyorgy1,Remaley Alan T

Affiliation:

1. Clinical Chemistry Service, Department of Laboratory Medicine, Warren Grant Magnuson Clinical Center, NIH, Bethesda, MD

Abstract

Abstract Background: Increased total triiodothyronine (TT3) assay results in apparently euthyroid patients triggered an investigation of the effect of blood collection tubes on serum TT3 and other laboratory assays. Methods: We examined potential assay interference for three types of tubes: plastic Greiner Bio-One™ Vacuette™; glass Becton Dickinson (BD) Vacutainer™; and plastic BD Vacutainer SST™ tubes. Serum samples from apparently healthy volunteers (age range, 30–60 years; 15 males and 34 females) were collected in different tube types and analyzed in 17 immunoassays (n = 49), 30 clinical chemistry tests (n = 20), and 33 immunology assays (n = 15). Tube effects were also examined by adding pooled serum to different tube types. Results: TT3 values, when measured by the IMMULITE™ 2000 but not the AxSYM™ analyzer, were significantly higher (P <0.0001) for SST (2.81 nmol/L) than either glass (2.15 nmol/L) or Vacuette (2.24 nmol/L) tubes. The effect was large enough to substantially shift the distribution of patient values, increasing the percentage of values above the reference interval from 11.3% to 35.8%. The degree of interference from SST tubes on TT3 differed among various tube lots and could be attributed to a tube additive shared by other plastic tubes. Results from several other tests statistically differed among tube types, but differences were not considered to be clinically significant. Conclusions: Assay interferences from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality-control or proficiency testing programs. Laboratories can, however, address this problem by monitoring distribution of patients’ results.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3