Affiliation:
1. Wisconsin State Laboratory of Hygiene, University of Wisconsin, Madison
Abstract
Abstract
We evaluated a plastic evacuated blood-drawing tube containing an integral serum-separating barrier gel, by direct comparison with a glass counterpart. The plastic tube demonstrated no differences when compared for common clinical chemistry analytes with multiple types of instruments and systems. A total of 260 such different combinations were studied with emphasis on tests sensitive to drawing and handling indexes such as lactate dehydrogenase and potassium. A total of six separate blood drawings were tested with no significant differences noted in these tests. The total study included subjective evaluations of the plastic tube's use as a blood-drawing device and objective studies based on quantitative test results from normal and hospitalized patients and use of the primary sampling tubes (both plastic and glass) for 48-h storage.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry