Abstract
Abstract
Background: Current clinical methods for folate give different results and cannot measure the various forms of folate. We developed an isotope-dilution tandem mass spectrometric method coupled to liquid chromatography (LC/MS/MS) as a candidate reference method for 5-methyltetrahydrofolic acid (5MeTHF), 5-formyltetrahydrofolic acid (5FoTHF), and folic acid (FA) in human serum.
Methods: We quantitatively isolated folates from 275 μL of serum with a phenyl solid-phase extraction cartridge, then detected and quantified them in stabilized serum extracts by positive-ion electrospray ionization LC/MS/MS. We used an isocratic mobile phase of acetic acid in organic solvent on a C8 analytical column. 13C-labeled folates were used as internal standards.
Results: Limits of detection in serum were 0.13 (5MeTHF), 0.05 (5FoTHF), and 0.07 (FA) nmol/L. Within- and between-run imprecision (CV) was <7% for 5MeTHF and <10% for 5FoTHF at concentrations >0.5 nmol/L, and <10% for FA at concentrations >2.0 nmol/L. Total folate (TFOL) concentrations determined by competitive protein binding radioassay were ∼9% lower than results obtained with LC/MS/MS. The microbiologic assay gave ∼15% higher TFOL results with FA calibrator and no difference with 5MeTHF calibrator. The mean (SD) [range] TFOL in 42 sera was 35.5 (17.8) [6.5–75.6] nmol/L. Thirty-two samples with TFOL <50 nmol/L had, on average, 93.3% 5MeTHF, 2.3% FA, and 4.4% 5FoTHF. Ten samples with TFOL >50 nmol/L had, on average, 81.7% 5MeTHF, 15.7% FA, and 2.5% 5FoTHF.
Conclusions: This stable-isotope-dilution LC/MS/MS method can quantify 5MeTHF, 5FoTHF, and FA in serum. Currently used clinical assays agree with this candidate reference method.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
205 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献