Background: External quality assessment schemes (EQAS) are conducted to evaluate user performance (participant assessment) and to assess different methods and instruments (method assessment). The quality of control materials is crucial to achieving these goals. Inconsistencies in between-lot variations detected by use of different control and sample materials may affect EQAS outcomes.
Methods: For the Accu-Chek Sensor, Precision Xtra, Ascensia Elite, and HemoCue 201 glucometers, 3 different lots of glucose strips were used with each instrument. Method assessment results from analysis of capillary blood and 3 control materials were used to calculate between-lot differences. A simulation study was performed to evaluate the effect of between-lot variation on participant assessment results.
Results: With the Precision Xtra, the results obtained with EQA control material mirrored those obtained with capillary blood, but for the other instruments, we found between-lot differences of as much as 1.3 mmol/L, which were substantially greater than those found with capillary blood and of clinical importance at decision limits. The simulation study showed an effect on participant assessment results related to the target values, with the percentage of poor results decreasing (38%, 10%, and 4%) with the use of common, method-specific, and lot-specific target values, respectively.
Conclusions: Between-lot variation may influence participant EQA results for participant and method assessments. The clinical relevance of between-lot variation discovered in EQAS using noncommutable control materials should be examined by use of native blood samples.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Reference25 articles.
1. Libeer JC, Baadenhuijsen H, Fraser CG, Petersen PH, Ricos C, Stockl D, et al. Characterization and classification of external quality assessment schemes (EQA) according to objectives such as evaluation of method and participant bias and standard deviation. External Quality Assessment (EQA) Working Group A on Analytical Goals in Laboratory Medicine. Eur J Clin Chem Clin Biochem1996;34:665-678.
2. Sciacovelli L, Zardo L, Secchiero S, Plebani M. Quality specifications in EQA schemes: from theory to practice. Clin Chim Acta2004;346:87-97.
3. ILAC Technical Accreditation Issues Committee. ILAC-G13:2000—guidelines for the requirements for the competence of providers for proficiency testing schemes. http://www.ilac.org (accessed January 2005)..
4. Thienpont LM, Leenheer AP, Stockl D, Reinauer H. Candidate reference methods for determining target values for cholesterol, creatinine, uric acid, and glucose in external quality assessment and internal accuracy control. II. Method transfer. Clin Chem1993;39:1001-1006.
5. Burrin J, Alberti K. What is blood glucose: can it be measured?. Diabetic Med1990;7:199-206.
Cited by
67 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献