Six of Eight Hemoglobin A1c Point-of-Care Instruments Do Not Meet the General Accepted Analytical Performance Criteria

Author:

Lenters-Westra Erna12,Slingerland Robbert J12

Affiliation:

1. Department of Clinical Chemistry, Isala klinieken, Zwolle, The Netherlands

2. European Reference Laboratory for Glycohemoglobin, Zwolle, The Netherlands

Abstract

Abstract Background: Hemoglobin A1c (Hb A1c) point-of-care (POC) instruments are widely used to provide rapid-turnaround results in diabetic care centers. We investigated the conformance of various Hb A1c POC instruments (In2it from Bio-Rad, DCA Vantage from Siemens, Afinion and Nycocard from Axis-Shield, Clover from Infopia, InnovaStar from DiaSys, A1CNow from Bayer, and Quo-Test from Quotient Diagnostics) with generally accepted performance criteria for Hb A1c. Methods: The CLSI protocols EP-10, EP-5, and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using 3 certified secondary reference measurement procedures and the mean of the 3 reference methods. Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria, as calculated from analyses with 2 different reagent lot numbers for each Hb A1c method, was also evaluated. Results: Because of disappointing EP-10 results, 2 of the 8 manufacturers decided not to continue the evaluation. The total CVs from EP-5 evaluations for the different instruments with a low and high Hb A1c value were: In2it 4.9% and 3.3%, DCA Vantage 1.8% and 3.7%, Clover 4.0% and 3.5%, InnovaStar 3.2% and 3.9%, Nycocard 4.8% and 5.2%, and Afinion 2.4% and 1.8%. Only the Afinion and the DCA Vantage passed the NGSP criteria with 2 different reagent lot numbers. Conclusions: Only the Afinion and the DCA Vantage met the acceptance criteria of having a total CV <3% in the clinically relevant range. The EP-9 results and the calculations of the NGSP certification showed significant differences in analytical performance between different reagent lot numbers for all Hb A1c POC instruments.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 195 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3