Abstract
Abstract
Background: Matrix effects can profoundly reduce the performance of electrospray ionization mass spectrometry. Preliminary observations indicated that the methanol used in the mobile phase could be a source of differential ionization or ion suppression.
Methods: Drug stability studies, analysis of biological extracts, mixing experiments, and postcolumn infusions were used to test 9 commercial methanols for ionization differences in liquid chromatography-tandem mass spectrometry assays for immunosuppressants. Area responses for the drugs and internal standards were compared for mobile phases prepared with each selected methanol. Postcolumn infusion experiments were performed to confirm the degree of ionization differences occurring at the ion source, and to evaluate the proportions of ammonium, sodium, and potassium adducts.
Results: The decrease in signal for the immunosuppressant drugs was shown to result from differential ionization associated with the selected methanols. Product ion intensity varied by 10-fold among the methanols tested. For sirolimus, tacrolimus, and mycophenolic acid, the percentage change in ionization was the same for the drug and its corresponding internal standard. Postcolumn sirolimus infusion evaluation revealed that a 1000-fold analyte concentration difference did not affect ionization. The proportions of ammonium, sodium, and potassium adducts of sirolimus precursor ions differed in relation to the source of methanol.
Conclusions: Organic solvents used in mobile phases and extract preparation of biological samples may be associated with ion suppression, affecting adduct formation and assay sensitivity.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
56 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献