Comparison of Point-of-Care Activated Clotting Time Methods in Different Clinical Settings in a Large Academic Medical Center

Author:

Korpi-Steiner Nichole L1,Walz J Matthias2,Schanzer Andres3,Rao Lokinendi V45

Affiliation:

1. Department of Pathology and Laboratory Medicine, University of North Carolina, Chapel Hill, NC

2. Department of Anesthesiology, UMass Memorial Medical Center, Worcester, MA

3. Department of Surgery, UMass Memorial Medical Center, Worcester, MA

4. Department of Pathology, UMass Memorial Medical Center, Worcester, MA

5. Quest Diagnostics, North Region, Marlborough, MA

Abstract

Abstract Objectives This study is a comparative analysis of measured activated clotting time (ACT) values by use of 5 different point-of-care (POC) ACT methods spanning the range detected during different clinical procedures at our institution. Methods We determined the correlation, imprecision, and differences in measured ACT values with use of 4 POC ACT methods compared with a reference ACT method in 41 venous whole blood samples collected from 25 adult patients undergoing interventional procedures. The POC ACT methods evaluated included the i-STAT with kaolin activator in prewarm mode, i-STAT with Celite activator in prewarm and nonprewarm modes, ACTPlus, and HMSPlus, which was designated the reference method. Each venous whole blood patient sample was tested in duplicate on each POC ACT test system (total n = 410 ACT measurements). Analyses of imprecision and differences in measured ACT values were stratified by moderate (100–299 s) and high (≥300 s) ACT ranges. Results In this study population, measured ACT values ranged from 100–835 s with use of the HMSPlus. All methods demonstrated good correlation (r ≥ 0.95) in ACT values compared to the reference method. Imprecision varied by method with ranges of 1.7%–2.7% CV in the moderate ACT range and 2.5%–4.8% CV in the high ACT range. ACTPlus and i-STAT-Celite-prewarm methods exhibited proportional differences in measured ACT values whereas the i-STAT-Celite-nonprewarm and i-STAT-kaolin-prewarm demonstrated constant differences in measured ACT values compared to HMSPlus. Conclusions ACT values correlate well between POC methods. Imprecision and difference profiles vary by method; notably, imprecision exceeds systematic differences in the high ACT range and contributes to intermethod differences that are limitations worthy of consideration when contemplating a change in ACT methods.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Reference17 articles.

1. Anticoagulation monitoring part 2: unfractionated heparin and low-molecular-weight heparin;Spinler;Ann Pharmacother,2005

2. Henry's clinical diagnosis and management by laboratory methods;McPherson,2011

3. Clinical evaluation of the i-STAT kaolin activated clotting time (ACT) test in different clinical settings in a large academic urban medical center: comparison with the Medtronic ACT Plus;Lewandrowski;Am J Clin Pathol,2011

4. Predictive performance of three methods of activated clotting time measurement in neonatal ECMO patients;Seay;ASAIO J,1993

5. Comparison of point-of-care activated clotting time systems utilized in a single pediatric institution;Ojito;J Extra Corpor Technol,2012