Application of a Simulation Model to Estimate Treatment Error and Clinical Risk Derived from Point-of-Care International Normalized Ratio Device Analytic Performance

Author:

Lyon Martha E1,Sinha Roona2,Lyon Oliver A S3,Lyon Andrew W1

Affiliation:

1. Department of Pathology and Laboratory Medicine, Division of Clinical Biochemistry, Saskatoon Health Region, Saskatoon, Saskatchewan, Canada

2. Department of Pediatrics, University of Saskatchewan, Saskatoon Health Region, Saskatoon, Saskatchewan, Canada

3. University of Saskatchewan, Saskatoon Health Region, Saskatoon, Saskatchewan, Canada

Abstract

Abstract Background In 2016, the Food and Drug Administration (FDA) proposed to enhance performance expectations for point-of-care testing (POCT) international normalized ratio (INR) devices relative to International Organization for Standardization (ISO) 17593:2007. The objective of the study was to estimate the frequency of warfarin dosing errors associated with a central laboratory INR method, a POCT INR method, and the proposed FDA performance goals. Methods A data set of INR results (n = 51912) from adult patients with INR ≤4 was used to assess the influence of adding assay imprecision and bias on warfarin dose decisions. The frequency of error in warfarin dose and size of error (≥1 or ≥2 dose categories) was compared using published assay specifications for the Instrumentation Laboratory ACL TOP® and the Roche Diagnostics CoaguChek® XS relative to the proposed FDA guidelines. Results The frequency of warfarin dose misclassification was largely influenced by bias and was not sensitive to assay imprecision. The central laboratory and POCT INR methods met the FDA performance specifications, had equal rates of ≥2 warfarin dose category error, and had statistically different rates of ≥1 warfarin dose category error in large samples (n >250). Conclusions Simulation models are useful tools for evaluating POCT INR assay performance criteria required to achieve the proposed FDA guidelines. This simulation depicted how the Roche Diagnostics CoaguChek XS instrument meets the guideline.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Reference21 articles.

1. American Heart Association/American College of Cardiology Foundation guide to warfarin therapy;Hirsh;Circulation,2003

2. Advantages of a warfarin protocol for long term care pharmacists: a retrospective cohort study;Sargent;Can Geriatr J,2016

3. Pharmacologic therapies in anticoagulation;Ferreira;Med Clin North Am,2016

4. Comparison of international normalized ration measurement between CoaguChek XS Plus and STA-R coagulation analyzers;Hur;Biomed Res Int,2013

5. Accuracy of capillary whole blood international normalized ration on the CoaguChek S, CoaguChek XS, and ISTAT 1 point-of-care analyzers;Karon;Am J Clin Pathol,2008