Cone Beam CT (CBCT) Evaluation of Inter- and Intra-Fraction Motion for Patients Undergoing Brain Radiotherapy Immobilized using a Commercial Thermoplastic Mask on a Robotic Couch

Author:

Lightstone A. W.1,Tsao May1,Basran Parminder S.2,Chan Gordon1,Pang Geordi1,Ma Lijun3,Lochray Fiona1,Sahgal Arjun1

Affiliation:

1. Department of Radiation Oncology, Odette Cancer Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, ON, Canada M4N 3M5

2. Department of Medical Physics, British Columbia Cancer Agency-Vancouver Island Centre, Victoria, BC, Canada V8R 6V5

3. Department of Radiation Oncology, University of California San Francisco, San Francisco, USA 94143-0226

Abstract

Patients receiving fractionated intensity-modulated radiation therapy (IMRT) for brain tumors are often immobilized with a thermoplastic mask; however, masks do not perfectly re-orient the patient due to factors including the maximum pressure which can be applied to the face, deformations of the mask assembly, patient compliance, etc. Consequently, ~3–5 mm PTV margins (beyond the CTV) are often recommended. We aimed to determine if smaller PTV margins are feasible using mask immobilization coupled with 1) a gantry mounted CBCT image guidance system and 2) position corrections provided by a full six-degree of freedom (6-DOF) robotic couch. A cohort of 34 brain tumor patients was treated with fractionated IMRT. After the mask set-up, an initial CBCT was obtained and registered to the planning CT. The robotic couch corrected the misalignments in all 6-DOF and a pre-treatment verification CBCT was then obtained. The results indicated a repositioning alignment within our threshold of 1.5 mm (3D). Treatment was subsequently delivered. A post-treatment CBCT was obtained to quantify intra-fraction motion. Initial, pre-treatment and post-treatment CBCT image data was analyzed. A total of 505 radiation fractions were delivered to the 34 patients resulting in ~1800 CBCT scans. The initial median 3D (magnitude) set-up positioning error was 2.60 mm. Robotic couch corrections reduced the 3D median error to 0.53 mm prior to treatment. Intra-fraction movement was responsible for increasing the median 3D positioning error to 0.86 mm, with 8% of fractions having a 3D positioning error greater than 2 mm. Clearly CBCT image guidance coupled with a robotic 6-DOF couch dramatically improved the positioning accuracy for patients immobilized in a thermoplastic mask system; however, such intra-fraction motion would be too large for single fraction radiosurgery.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3