Author:
Lau Lawrence Chun Man,Lee Wayne Yuk Wai,Butler Anthony P. H.,Chernoglazov Alex I.,Chung Kwong Yin,Ho Kevin Ki Wai,Griffith James,Butler Philip H.,Yung Patrick Shu Hang
Abstract
AbstractTo determine whether state-of-the-art multi-energy spectral photon-counting computed tomography (MARS) can detect knee arthroplasty implant failure not detected by standard pre-operative imaging techniques. A total knee arthroplasty (TKA) removed from a patient was reviewed. The extracted prosthesis [NexGen Legacy Posterior Stabilized (LPS) TKA] was analyzed as were pre-operative imaging examination and compared with a MARS-CT examination obtained of the extracted TKA prosthesis. Radiographs, fluoroscopy, ultrasound and MRI preoperatively did not reveal the cause of the implant failure. MARS CT images of the extracted prosthesis clearly showed the presence of posteromedial polyethylene and tibial tray wear which is compatible with the clinical appearance of the extracted TKA. MARS can identify polyethylene insert and metallic tibial tray wear as a cause of TKA failure, that could not be identified with on standard pre-operative imaging. Although clinical MARS CT system is still under development, this case does illustrate its potential clinical usefulness. This is the first study to document how MARS CT imaging can detect orthopedic implant failure not detected by standard current imaging techniques. This system has a potential clinical application in orthopedic patients.
Publisher
Springer Science and Business Media LLC
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献