Abstract
Abstract
Background
Successful osseointegration of joint replacement implants is required for long-term implant survival. Accurate assessment of osseointegration could enable clinical discrimination of failed implants from other sources of pain avoiding unnecessary surgeries. Photon-counting detector computed tomography (PCD-CT) provides improvements in image resolution compared to conventional energy-integrating detector CT (EID-CT), possibly allowing better visualization of bone-implant-interfaces and osseointegration. The aim of this study was to assess the quality of visualization of bone-implant-interfaces and osseointegration in acetabular cup implants, using PCD-CT compared with EID-CT.
Methods
Two acetabular implants (one cemented, one uncemented) retrieved during revision surgery were scanned using PCD-CT and EID-CT at equal radiation dose. Images were reconstructed using different reconstruction kernels and iterative strengths. Delineation of the bone-implant and bone-cement-interface as an indicator of osseointegration was scored subjectively for image quality by four radiologists on a Likert scale and assessed quantitatively.
Results
Delineation of bone-implant and bone-cement-interfaces was better with PCD-CT compared with EID-CT (p ≤ 0.030). The highest ratings were given for PCD-CT at sharper kernels for the cemented cup (PCD-CT, median 5, interquartile range 4.25–5.00 versus EID-CT, 3, 2.00–3.75, p < 0.001) and the uncemented cup (5, 4.00–5.00 versus 2, 2–2, respectively, p < 0.001). The bone-implant-interface was 35–42% sharper and the bone-cement-interface was 28–43% sharper with PCD-CT compared with EID-CT, depending on the reconstruction kernel.
Conclusions
PCD-CT might enable a more accurate assessment of osseointegration of orthopedic joint replacement implants.
Key points
• The bone-implant interface ex vivo showed superior visualization using photon-counting detector computed tomography (PCD-CT) compared to energy-integrating detector computed tomography.
• Harder reconstruction kernels in PCD-CT provide sharper images with lower noise levels.
• These improvements in imaging might make it possible to visualize osseointegration in vivo.
Funder
Academy Medical Science Fund, Royal Netherlands Academy of Arts & Sciences
ALF grants Östergötland, Sweden
Knut och Alice Wallenbergs Stiftelse
Linköping University
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献