Abstract
AbstractLong-term bioelectronic implants require stable, hermetic encapsulation. Water and ion ingress are challenging to quantify, especially in miniaturized microsystems and over time. We propose a wireless and battery-free flexible platform leveraging backscatter communication and magnesium (Mg)-based microsensors. Water permeation through the encapsulation induces corrosion of the Mg resistive sensor thereby shifting the oscillation frequency of the sensing circuit. Experimental in vitro and in-tissue characterization provides information on the operation of the platform and demonstrates the robustness and accuracy of this promising method, revealing its significance for in-situ real-time monitoring of implanted bioelectronics.
Publisher
Springer Science and Business Media LLC