Author:
Clare Kevin,Stein Alan,Damodara Nitesh,Feldstein Eric,Alshammari Hussein,Ali Syed,Kurian Christeena,Rosenberg Jon,Bauerschmidt Andrew,Kaur Gurmeen,Santarelli Justin,Hamilton Robert,Mayer Stephan,Gandhi Chirag D.,Al-Mufti Fawaz
Abstract
AbstractDelayed cerebral ischemia (DCI) secondary to vasospasm is a determinate of outcomes following non-traumatic subarachnoid hemorrhage (SAH). SAH patients are monitored using transcranial doppler (TCD) to measure cerebral blood flow velocities (CBFv). However, the accuracy and precision of manually acquired TCD can be operator dependent. The NovaGuide robotic TCD system attempts to standardize acquisition. This investigation evaluated the safety and efficacy of the NovaGuide system in SAH patients in a Neuro ICU. We retrospectively identified 48 NovaGuide scans conducted on SAH patients. Mean and maximum middle cerebral artery (MCA) CBFv were obtained from the NovaGuide and the level of agreement between CBFv and computed tomography angiography (CTA) for vasospasm was determined. Safety of NovaGuide acquisition of CBFv was evaluated based on number of complications with central venous lines (CVL) and external ventricular drains (EVD). There was significant agreement between the NovaGuide and CTA (Cohen’s Kappa = 0.74) when maximum MCA CBFv ≥ 120 cm/s was the threshold for vasospasm. 27/48 scans were carried out with CVLs and EVDs present without negative outcomes. The lack of adverse events associated with EVDs/CVLs and the strong congruence between maximal MCA CBFv and CTA illustrates the diagnostic utility of the NovaGuide.