Author:
Maubert Bertrand,Theillière Camille,Jany Prescillia,Bourlet Thomas,Deschamps Jérôme,Pozzetto Bruno,Singh Fateh,Gadea Emilie
Abstract
AbstractCovid-19 has spurred a renewed interest in decontamination techniques for air, objects and surfaces. Beginning in 2020, urgent effort was done to permit the reuse of UV-C for inactivating SARS-CoV-2. However, those studies diverged widely on the dose necessary to reach this goal; until today, the real value of the sensitivity of the virus to a 254-nm illumination is not known precisely. In this study, decontamination was performed in an original UV-C large decontamination chamber (UVCab, ON-LIGHT, France) delivering an omnidirectional irradiation with an average dose of 50 mJ/cm2 in 60 s. Viral inactivation was checked by both cell culture and PCR test. SARS-CoV-2 was inactivated by UV-C light within 3 s on both porous (disposable gown) and non-porous (stainless steel and apron) surfaces. For the porous surface, an irradiation of 5 min was needed to achieve a completely negative PCR signal. The Z value estimating the sensitivity of SARS-CoV-2 to UV-C in the experimental conditions of our cabinet was shown to be > 0.5820 m2/J. These results illustrate the ability of this apparatus to inactivate rapidly and definitively high loads of SARS-CoV-2 deposited on porous or non-porous supports and opens new perspectives on material decontamination using UV-C.
Publisher
Springer Science and Business Media LLC