Author:
Alharthi Sarah,Ali Ashraf,Iqbal Muzaffar,Ibrar Aliya,Ahmad Bashir,Nisa Sobia,Mabood Fazal
Abstract
AbstractPorous silica particles were prepared by sol–gel method with some modification to get wide-pore particles. These particles were derivatized with N-phenylmaleimide-methylvinylisocyanate (PMI) and styrene by reversible addition fragmentation chain transfer (RAFT) polymerization to prepare N-phenylmaleimide embedded polystyrene (PMP) stationary phases. Narrow bore stainless steel column (100 × 1.8 mm i.d) was packed by slurry packing method. The chromatographic performance of PMP column was evaluated for the separation of synthetic peptides mixture composed of five peptides (Gly-Tyr, Gly-Leu-Tyr, Gly-Gly-Tyr-Arg, Tyr-Ile-Gly-Ser-Arg, Leucine enkephalin) and tryptic digest of human serum albumin (HAS) respectively. Number of theoretical plates as high as 280,000 plates/m were obtained for peptides mixture at optimum elution condition. Separation performance of the developed column was compared with commercial Ascentis Express RP-Amide column and it was observed that separation performance of PMP column was better than commercial column in terms of separation efficiency and resolution.
Publisher
Springer Science and Business Media LLC
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献