Author:
Tsuchida Takumi,Hayakawa Mineji,Kumano Osamu
Abstract
AbstractGlobal fibrinolysis assays detect the fibrinolysis time of clot dissolution using tissue-type plasminogen activator (tPA). Two such assays, clot-fibrinolysis waveform analysis (CFWA) and global fibrinolysis capacity (GFC) assay, were recently developed. These were compared with rotational thromboelastography (ROTEM). Healthy donor blood samples were divided into four groups based on tPA-spiked concentrations: 0, 100, 500, and 1000 ng/mL. CFWA and GFC fibrinolysis times, including 4.1 µg/mL and 100 ng/mL tPA in the assays, were determined, denoted as CFWA-Lys and GFC-Lys, respectively. Statistical differences were recognized between tPA concentrations of 0 and 500/1000 ng/mL for CFWA-Lys, and 0 and 100/500/1000 ng/mL for GFC-Lys. The correlation coefficients with lysis onset time (LOT) of extrinsic pathway evaluation and intrinsic pathway evaluation in ROTEM were statistically significant at 0.610 and 0.590 for CFWA-Lys, and 0.939 and 0.928 for GFC-Lys, respectively (p-values < 0.0001 for all correlations). Both assays showed significant correlations with ROTEM; however, the GFC assay proved to have better agreement with ROTEM compared with the CFWA assay. These assays have the potential to reflect a hyperfibrinolysis status with high tPA concentrations.