Feasibility of Magnetic Resonance–Compatible Accelerometers to Monitor Tremor Fluctuations During Magnetic Resonance–Guided Focused Ultrasound Thalamotomy: Technical Note

Author:

Baek Hongchae12ORCID,Chen Jacqueline1,Lockwood Daniel1,Obusez Emmanuel1,Poturalski Matthew1,Nagel Sean J.2,Jones Stephen E.1

Affiliation:

1. Imaging Institute, Cleveland Clinic, Cleveland, Ohio, USA;

2. Center for Neurological Restoration, Neurological Institute, Cleveland Clinic, Cleveland, Ohio, USA

Abstract

BACKGROUND: Magnetic resonance–guided focused ultrasound (MRgFUS) thalamotomy is used to treat essential tremor and tremor-dominant Parkinson disease. Feedback is collected throughout the procedure to verify the location of the target and completeness of response; however, variability in clinical judgments may underestimate or overestimate treatment response. OBJECTIVE: To objectively quantify joint motion after each sonication using accelerometers secured to the contralateral upper extremity in an effort to optimize MRgFUS treatment. METHODS: Before the procedure, 3 accelerometers were secured to the patient’s arm, forearm, and index finger. Throughout the procedure, tremor motion was regularly recorded during postural and kinetic tremor testing and individual joint angle measures were modeled. The joint angle from each accelerometer was compared with baseline measurements to assess changes in angles. Subsequent adjustments to the target location and sonication energy were made at the discretion of the neurosurgeon and neuroradiologist. RESULTS: Intraoperative accelerometer measurements of hand tremor from 18 patients provided quantified data regarding joint angle reduction: 87.3%, 94.2%, and 86.7% for signature writing, spiral drawing, and line drawing tests, respectively. Target adjustment based on accelerometer monitoring of the angle at each joint added substantial value toward achieving optimal tremor reduction. CONCLUSION: Real-time accelerometer recordings collected during MRgFUS thalamotomy offered objective quantification of changes in joint angle after each sonication, and these findings were consistent with clinical judgments of tremor response. These results suggest that this technique could be used for fine adjustment of the location of sonication energy and number of sonications to consistently achieve optimal tremor reduction.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3