van Brummelen Emilie M. J.,Ziagkos Dimitrios,de Boon Wadim M. I.,Hart Ellen P.,Doll Robert J.,Huttunen Teppo,Kolehmainen Petteri,Groeneveld Geert Jan
Abstract
Abstract
Background
To quantify pharmacological effects on tremor in patients with essential tremor (ET) or Parkinson’s Disease (PD), laboratory-grade accelerometers have previously been used. Over the last years, consumer products such as smartphones and smartwatches have been increasingly applied to measure tremor in an easy way. However, it is unknown how the technical performance of these consumer product accelerometers (CPAs) compares to laboratory-grade accelerometers (LGA). This study was performed to compare the technical performance of CPAs with LGA to measure tremor in patients with Parkinson’s Disease (PD) and essential tremor (ET).
Methods
In ten patients with PD and ten with ET, tremor peak frequency and corresponding amplitude were measured with 7 different CPAs (Apple iPhone 7, Apple iPod Touch 5, Apple watch 2, Huawei Nexus 6P, Huawei watch, mbientlabMetaWear (MW) watch, mbientlab MW clip) and compared to a LGA (Biometrics ACL300) in resting and extended arm position.
Results
Both in PD and ET patients, the peak frequency of CPAs did not significantly differ from the LGA in terms of limits of agreement. For the amplitude at peak frequency, only the iPhone and MW watch performed comparable to the LGA in ET patients, while in PD patients all methods performed comparable except for the iPod Touch and Huawei Nexus. Amplitude was higher when measured with distally-located CPAs (Clip, iPhone, iPod) compared with proximally-located CPAs (all watches). The variability between subjects was higher than within subjects for frequency (25.1% vs. 13.4%) and amplitude measurement (331% vs. 53.6%). Resting arm position resulted in lower intra-individual variability for frequency and amplitude (13.4 and 53.5%) compared to extended arm position (17.8 and 58.1%).
Conclusions
Peak frequencies of tremor could be measured with all tested CPAs, with similar performance as LGA. The amplitude measurements appeared to be driven by anatomical location of the device and can therefore not be compared. Our results show that the tested consumer products can be used for tremography, allowing at-home measurements, in particular in studies with a cross-over or intra-individual comparison design using the resting arm position.
Trial registration
This trial was registered in the Dutch Competent Authority (CCMO) database with number NL60672.058.17 on May 30th 2017.