High-resolution tomographic imaging of a human cerebellum: comparison of absorption and grating-based phase contrast

Author:

Schulz Georg1,Weitkamp Timm2,Zanette Irene2,Pfeiffer Franz3,Beckmann Felix4,David Christian5,Rutishauser Simon5,Reznikova Elena6,Müller Bert1

Affiliation:

1. Biomaterials Science Center, University of Basel, Basel, Switzerland

2. European Synchrotron Radiation Facility, Grenoble, France

3. Department of Physics/Biophysics (E17), Technische Universität München, Garching, Germany

4. Institute for Materials Research, GKSS Research Centre, Geesthacht, Germany

5. Laboratory for Micro- and Nanotechnology, Paul Scherrer Institut, Villigen, Switzerland

6. Institute of Microstructure Technology, Karlsruhe Institute for Technology, Karlsruhe, Germany

Abstract

Human brain tissue belongs to the most impressive and delicate three-dimensional structures in nature. Its outstanding functional importance in the organism implies a strong need for brain imaging modalities. Although magnetic resonance imaging provides deep insights, its spatial resolution is insufficient to study the structure on the level of individual cells. Therefore, our knowledge of brain microstructure currently relies on two-dimensional techniques, optical and electron microscopy, which generally require severe preparation procedures including sectioning and staining. X-ray absorption microtomography yields the necessary spatial resolution, but since the composition of the different types of brain tissue is similar, the images show only marginal contrast. An alternative to absorption could be X-ray phase contrast, which is known for much better discrimination of soft tissues but requires more intricate machinery. In the present communication, we report an evaluation of the recently developed X-ray grating interferometry technique, applied to obtain phase-contrast as well as absorption-contrast synchrotron radiation-based microtomography of human cerebellum. The results are quantitatively compared with synchrotron radiation-based microtomography in optimized absorption-contrast mode. It is demonstrated that grating interferometry allows identifying besides the blood vessels, the stratum moleculare, the stratum granulosum and the white matter. Along the periphery of the stratum granulosum, we have detected microstructures about 40 µm in diameter, which we associate with the Purkinje cells because of their location, size, shape and density. The detection of individual Purkinje cells without the application of any stain or contrast agent is unique in the field of computed tomography and sets new standards in non-destructive three-dimensional imaging.

Publisher

The Royal Society

Subject

Biomedical Engineering,Biochemistry,Biomaterials,Bioengineering,Biophysics,Biotechnology

Reference51 articles.

1. Micromechanical response of mineral and collagen phases in bone

2. Non-rigid registration of multi-modal images using both mutual information and cross-correlation

3. A novel system for X-ray phase-contrast microtomography;Beckmann F.;Hasylab Annu. Rep.,1995

4. Microtomography using synchrotron radiation at DESY: current status and future developments

5. High density resolution in synchrotron-radiation-based attenuation-contrast microtomography;Beckmann F.;Proc. SPIE,2008