Affiliation:
1. Istanbul Medeniyet University, Turkey
2. Sisli Hamidiye Etfal Training and Research Hospital, Turkey
3. Umraniye Training and Research Hospital, Turkey
Abstract
ABSTRACT Objective: To compare antirotator proximal femoral nail (A-PFN) with antirotator dynamic hip screw (A-DHS). Methods: Fourteen proximal femur models with type 31/A2 fracture, according to the AO Foundation/Orthopaedic Trauma Association (AO/OTA) classification, were separated into two groups. Group 1 bones (n = 7) were fixed with A-PFN and Group 2 (n = 7) with A-DHS. A 5 mm/min axial load was applied to femur heads using a testing device. Results: Two of the seven models in the A-PFN group fractured at the proximal, and the other five at distal locking screw level. All models in the A-DHS group fractured at the tightened distal screw region. The median fracture load for the A-PFN group was 132.1 N (97.1-173.69 N range), and for the A-DHS group it was 81.7 N (75.15-89.12 N range). Conclusion: A-PFN-treated unstable intertrochanteric fractured models resisted to higher levels of axial load than the A-DHS-treated group, with statistically significant difference. However, clinical studies are required to support these results. Level of Evidence V, Biomechanical study.
Subject
Rehabilitation,Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献