Affiliation:
1. 1Department of Microbiology and Hospital Hygiene, Bundeswehr Hospital Hamburg, Hamburg, Germany
2. 2Institute for Medical Microbiology, Virology and Hygiene, University Medicine Rostock, Rostock, Germany
Abstract
AbstractIntroductionThe aim of the study was a comparative evaluation of in-house real-time PCR and commercial real-time PCR (Fast Track Diagnostics (FTD), ampliCube/Mikrogen) targeting enteropathogenic bacteria from stool in preparation of Regulation (EU) 2017/746 on in vitro diagnostic medical devices.MethodsBoth 241 stool samples from patients and 100 samples from German laboratory control schemes (“Ringversuche”) were used to comparatively assess in-house real-time PCR, the FTD bacterial gastroenteritis kit, and the ampliCube gastrointestinal bacterial panels 1&2 either with the in-house PCRs as gold standard and as a test comparison without gold standard applying latent class analysis. Sensitivity, specificity, intra- and inter-assay variation and Cohen’s kappa were assessed.ResultsIn comparison with the gold standard, sensitivity was 75–100% for strongly positive samples, 20–100% for weakly positive samples, and specificity ranged from 96 to 100%. Latent class analysis suggested that sensitivity ranges from 81.2 to 100% and specificity from 58.5 to 100%. Cohen’s kappa varied between moderate and nearly perfect agreement, intra- and inter-assay variation was 1–3 to 1–4 Ct values.ConclusionAcceptable agreement and performance characteristics suggested replaceability of the in-house PCR assays by the commercial approaches.
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献