Abstract
Background: In the literature no information about plates for the high-neck mandibular condylar osteosynthesis could be found despite that 30 plate designs have been published. The main course consider the basal condylar or diacapitular fractures. The aim of the study was to test mechanically all available designs (only 4 of 30 was proper) on polyurethane mandibles using an individually designed clamping system. Methods: Forces required for a 1 mm displacement of fixed fracture fragments and incidents of screw loosening were recorded. Results: It has occured that dedicated plates for fixation are much weaker than set of two straight plates (p < 0.0001). General observation is the bigger plate and more screws, the better rigid stable osteosynthesis of mandibular condyle, however, there are limitations in plates design for high-neck fractures resulted in restricted operation field. Conclusion: Double straight plates occured to be the best mechanical fixation for high-neck fractures of the mandibular condyle. Maybe other existing plates could be used but only after prebending or that fracture required novel dedicated plates design.
Funder
Uniwersytet Medyczny w Lodzi
Subject
General Materials Science
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献