Author:
Zhu Linlin,Dong Yuchu,Li Zexiao,Zhang Xiaodong
Abstract
The two peaks characteristic of yellow and blue light in the spectrum of dual-wavelength white light emitting diodes (LEDs) introduce distinctive features to the interference signal of white light scanning interferometry (WLSI). The distinctive features are defined as discontinuities, so that the fringe contrast function cannot be modeled as a single Gaussian function, and causes the interferogram to have uneven distribution of fringes of different orders in the scanning interferometer. This phenomenon leads to the low accuracy of the zero-order fringe position in the envelope calculation, which affects the repeatability and accuracy of the interferometry. This paper proposes a new surface recovery algorithm based on the Hilbert phase envelope and adjacent reference points calculation, which can effectively overcome the influence of the discontinuous signal of dual-wavelength LED white light interference on the three-dimensional reconstruction of WLSI measurements. The reliability of the algorithm is verified by experiments, and the measurement accuracy of LED WLSI system is evaluated.
Funder
National Key Research and Development Program of China
Science Challenge Program
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献