Abstract
Measuring cardiac activity from the chest using an accelerometer is commonly referred to as seismocardiography. Unfortunately, it cannot provide clinically valid data because it is easily corrupted by motion artefacts. This paper proposes two methods to improve peak detection from noisy seismocardiography data. They rely on discrete wavelet transform analysis using either biorthogonal 3.9 or reverse biorthogonal 3.9. The first method involves slicing chest vibrations for each cardiac activity, and then detecting the peak location, whereas the other method aims at detecting the peak directly from chest vibrations without segmentation. Performance evaluations were conducted on signals recorded from small children and adults based on missing and additional peaks. Both algorithms showed a low error rate (15.4% and 2.1% for children/infants and adults, respectively) for signals obtained in resting state. The average error for sitting and breathing tasks (adults only) was 14.4%. In summary, the first algorithm proved more promising for further exploration.
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献