Abstract
Wearable electrocardiogram (ECG) devices are universally used around the world for patients who have cardiovascular disease (CVD). At present, how to suppress motion artifacts is one of the most challenging issues in the field of physiological signal processing. In this paper, we propose an adaptive cancellation algorithm based on multi-inertial sensors to suppress motion artifacts in ambulatory ECGs. Firstly, this method collects information related to the electrode motion through multi-inertial sensors. Then, the part that is not related to the electrode motion is removed through wavelet transform, which improves the correlation of the reference input signal. In this way, the ability of the adaptive cancellation algorithm to remove motion artifacts is improved in the ambulatory ECG. Subsequent experimentation demonstrated that the wavelet adaptive cancellation algorithm based on multi-inertial sensors can effectively remove motion artifacts in ambulatory ECGs.
Funder
National Natural Science Foundation of China
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献