Abstract
The electrocardiogram (ECG) is one of the most commonly measured biosignals. In particular, textile electrodes allow for the measuring of long-term ECG without skin irritation or other discomforts for the patient. Such textile electrodes, however, usually suffer from insufficient or unreliable skin contact. Thus, developing textile electrodes is impeded by the often-complicated differentiation between signal artifacts due to moving and breathing and artifacts related to unreliable skin contact. Here, we suggest a simple method of using 50/60 Hz power grid noise to evaluate the skin contact of different textile electrodes in comparison with commercial glued electrodes. We use this method to show the drying of wetted skin under an embroidered electrode as well as sweating of the originally dry skin under a coated electrode with high water vapor resistance.
Funder
Federal Ministry of Education and Research
Reference49 articles.
1. The Leading Causes of Death in the US for 2020
2. Causes of Death;Ritchie;Our World in Data,2018
3. Prevalence and magnitude of ST-segment and T wave abnormalities in men during continuous ambulatory electrocardiography;Armstrong;Am. J. Cardiol.,1981
4. Development of Underwear with Integrated 12 Channel ECG for Men and Women
5. Integrating wireless ECG monitoring in textiles
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献