Abstract
This study aimed to analyze the influence of multiple uses of zirconia implant drills on their cutting performance and bending strength. The hypothesis was that drill usage and sterilization cycles would not affect drilling time or flexural strength. Sixty zirconia twist drills from Z-Systems were used to drill in the angulus mandibulae region of fresh porcine jaws. The drills were divided into four groups based on the cycle count, and the drilling time was measured. Bending strength tests were conducted using a universal testing machine, and statistical analysis was performed using ANOVA tests. The results showed that drilling times followed a normal distribution, and significant differences were observed in drilling times between group 1 and the other groups for the pilot drill. However, no significant differences were found for ø3.75 mm and ø4.25 mm drills, and drilling times also varied significantly among different drill diameters, regardless of the cycle count. Flexural strength did not significantly differ among drill diameters or sterilization cycles. Overall, using and sterilizing zirconia implant drills had no significant impact on drilling time or flexural strength. Nevertheless, drilling times did vary depending on the diameter of the drill. These findings provide valuable insights into the performance and durability of zirconia implant drills, contributing to the optimization of dental implant procedures.