Abstract
Current methods for measuring the abundance of proteogenic amino acids in plants require derivatisation, extended run times, very sensitive pH adjustments of the protein hydrolysates, and the use of buffers in the chromatographic phases. Here, we describe a fast liquid chromatography–mass spectrometry (LC–MS) method for the determination of amino acids that requires only three steps: hydrolysis, neutralisation, and sample dilution with a borate buffer solution for pH and retention time stability. The method shows excellent repeatability (repeated consecutive injections) and reproducibility (repeated hydrolysis) in the amino acid content, peak area, and retention time for all the standard amino acids. The chromatographic run time is 20 min with a reproducibility and repeatability of <1% for the retention time and <11% for the peak area of the BSA and quality control (QC) lentil samples. The reproducibility of the total protein levels in the hydrolysis batches 1–4 was <12% for the BSA and the lentil samples. The level of detection on column was below 0.1 µM for most amino acids (mean 0.017 µM).
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献