Measurement of Electrical Impedance Tomography-Based Regional Ventilation Delay for Individualized Titration of End-Expiratory Pressure

Author:

Muders ThomasORCID,Hentze Benjamin,Kreyer Stefan,Wodack Karin Henriette,Leonhardt SteffenORCID,Hedenstierna Göran,Wrigge Hermann,Putensen Christian

Abstract

Rationale: Individualized positive end-expiratory pressure (PEEP) titration might be beneficial in preventing tidal recruitment. To detect tidal recruitment by electrical impedance tomography (EIT), the time disparity between the regional ventilation curves (regional ventilation delay inhomogeneity [RVDI]) can be measured during controlled mechanical ventilation when applying a slow inflation of 12 mL/kg of body weight (BW). However, repeated large slow inflations may result in high end-inspiratory pressure (PEI), which might limit the clinical applicability of this method. We hypothesized that PEEP levels that minimize tidal recruitment can also be derived from EIT-based RVDI through the use of reduced slow inflation volumes. Methods: Decremental PEEP trials were performed in 15 lung-injured pigs. The PEEP level that minimized tidal recruitment was estimated from EIT-based RVDI measurement during slow inflations of 12, 9, 7.5, or 6 mL/kg BW. We compared RVDI and PEI values resulting from different slow inflation volumes and estimated individualized PEEP levels. Results: RVDI values from slow inflations of 12 and 9 mL/kg BW showed excellent linear correlation (R2 = 0.87, p < 0.001). Correlations decreased for RVDI values from inflations of 7.5 (R2 = 0.68, p < 0.001) and 6 (R2 = 0.42, p < 0.001) mL/kg BW. Individualized PEEP levels estimated from 12 and 9 mL/kg BW were comparable (bias −0.3 cm H2O ± 1.2 cm H2O). Bias and scatter increased with further reduction in slow inflation volumes (for 7.5 mL/kg BW, bias 0 ± 3.2 cm H2O; for 6 mL/kg BW, bias 1.2 ± 4.0 cm H2O). PEI resulting from 9 mL/kg BW inflations were comparable with PEI during regular tidal volumes. Conclusions: PEEP titration to minimize tidal recruitment can be individualized according to EIT-based measurement of the time disparity of regional ventilation courses during slow inflations with low inflation volumes. This sufficiently decreases PEI and may reduce potential clinical risks.

Funder

Deutsche Forschungsgemeinschaft

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3