Abstract
We assessed xurography and laser ablation for the manufacture of passive micromixers arrays to explore the scalability of unconventional manufacture technologies that could be implemented under the restrictions of the Point of Care for developing countries. In this work, we present a novel split-and-recombine (SAR) array design adapted for interfacing standardized dispensing (handheld micropipette) and sampling (microplate reader) equipment. The design was patterned and sealed from A4 sized vinyl sheets (polyvinyl chloride), employing low-cost disposable materials. Manufacture was evaluated measuring the dimensional error with stereoscopic and confocal microscopy. The micromixing efficiency was estimated using a machine vision system for passive driven infusion provided by micropippetting samples of dye and water. It was possible to employ rapid fabrication based on xurography to develop a four channel asymmetric split-and-recombine (ASAR) micromixer with mixing efficiencies ranging from 43% to 65%.
Funder
Consejo Nacional de Ciencia y Tecnología
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献