Abstract
Photoacoustic (PA) measurements with open resonators usually provide poor detection sensitivity due to signal leakage at the resonator opening. We have recently demonstrated three different approaches for modelling the photoacoustic signal of open resonators. In this work, one of the approaches is applied for the optimization of the geometry of the T-shaped resonator for improved signal strength and thus sensitivity. The results from the numerical optimization show an increase in the photoacoustic signal by a factor of approximately 7.23. They are confirmed using numerical methods other than the one applied for the optimization and by experimental measurement. The measurement shows an increase in the photoacoustic signal by a factor of approximately 2.34.
Funder
Hamburg University of Applied Science Promotionsförderung
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献