This ex vivo study aimed to measure the performance of an electronic apex locator (EAL) in the presence of sodium hypochlorite irrigants with different concentrations. Two EALs (Root ZX Mini and Locapex 6) were used to locate the apical foramen in 10 extracted single-rooted teeth in the presence of 0.5%, 2.5% and 5% sodium hypochlorite. Visual working lengths were also determined using #10 K-file under a microscope before the electronic measurements were made. The performance of both EALs was compared for the electronic working lengths determined under the different concentrations of sodium hypochlorite. A multiple-way ANOVA and PLSD Fisher’s test with an α risk fixed at 5% were conducted. There were no statistical differences in the working lengths determined by both EALs between the three groups with different concentrations of sodium hypochlorite and their visual control measurements. When a ± 0.5 mm margin was applied, the Root ZX Mini and the Locapex 6 presented 88% and 83% accuracy, respectively. Sodium hypochlorite concentration in irrigants does not affect the accuracy and reliability of either the Root ZX Mini or the Locapex 6. Electronic apex locators are reliable with any concentration of sodium hypochlorite irrigants.

Publisher

MDPI AG

Subject

General Materials Science

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3