Abstract
The paper presents the results of the investigations on the direction-dependent accuracy of the point identification during contact probe measurements with a coordinate measuring machine (CMM). Considering the contact point identified by an orthogonal to the surface probe movement, the transformation of coordinates was made in order to calculate the displacement of the measured point. As a result, the positioning accuracy was estimated in three axes. The experiments demonstrated a strong dependence of the displacement on the declination angle. Moreover, it was found that the directional surface texture which provided different roughness in perpendicular directions, had an impact on the positioning accuracy.
Funder
Ministry of Education, Youth and Sports
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry