Abstract
The purpose of this study was to assess the image quality of the low dose 2D/3D slot scanner (LDSS) imaging system compared to conventional digital radiography (DR) imaging systems. Visual image quality was assessed using the visual grading analysis (VGA) method. This method is a subjective approach that uses a human observer to evaluate and optimise radiographic images for different imaging technologies. Methods and materials: ten posterior-anterior (PA) and ten lateral (LAT) images of a chest anthropomorphic phantoms and a knee phantom were acquired by an LDSS imaging system and two conventional DR imaging systems. The images were shown in random order to three (chest) radiologists and three experienced (knee) radiographers, who scored the images against a number of criteria. Inter- and intraobserver agreement was assessed using Fleiss’ kappa and weighted kappa. Results: the statistical comparison of the agreement between the observers showed good interobserver agreement, with Fleiss’ kappa coefficients of 0.27–0.63 and 0.23–0.45 for the chest and knee protocols, respectively. Comparison of intraobserver agreement also showed good agreement with weighted kappa coefficients of 0.27–0.63 and 0.23–0.45 for the chest and knee protocols, respectively. The LDSS imaging system achieved significantly higher VGA image quality compared to the DR imaging systems in the AP and LAT chest protocols (p < 0.001). However, the LDSS imaging system achieved lower image quality than one DR system (p ≤ 0.016) and equivalent image quality to the other DR systems (p ≤ 0.27) in the knee protocol. The LDSS imaging system achieved effective dose savings of 33–52% for the chest protocol and 30–35% for the knee protocol compared with DR systems. Conclusions: this work has shown that the LDSS imaging system has the potential to acquire chest and knee images at diagnostic quality and at a lower effective dose than DR systems.
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献