Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor

Author:

Pei Zheyuan1ORCID,Balitskiy Maxim1,Thalman Ryan2ORCID,Kelly Kerry E.1

Affiliation:

1. Department of Chemical Engineering, University of Utah, Salt Lake City, UT 84112, USA

2. Department of Chemistry, Snow College, Ephraim, UT 84627, USA

Abstract

Formaldehyde is a known human carcinogen and an important indoor and outdoor air pollutant. However, current strategies for formaldehyde measurement, such as chromatographic and optical techniques, are expensive and labor intensive. Low-cost gas sensors have been emerging to provide effective measurement of air pollutants. In this study, we evaluated eight low-cost electrochemical formaldehyde sensors (SFA30, Sensirion®, Staefa, Switzerland) in the laboratory with a broadband cavity-enhanced absorption spectroscopy as the reference instrument. As a group, the sensors exhibited good linearity of response (R2 > 0.95), low limit of detection (11.3 ± 2.07 ppb), good accuracy (3.96 ± 0.33 ppb and 6.2 ± 0.3% N), acceptable repeatability (3.46% averaged coefficient of variation), reasonably fast response (131–439 s) and moderate inter-sensor variability (0.551 intraclass correlation coefficient) over the formaldehyde concentration range of 0–76 ppb. We also systematically investigated the effects of temperature and relative humidity on sensor response, and the results showed that formaldehyde concentration was the most important contributor to sensor response, followed by temperature, and relative humidity. The results suggest the feasibility of using this low-cost electrochemical sensor to measure formaldehyde concentrations at relevant concentration ranges in indoor and outdoor environments.

Funder

National Science Foundation

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Reference61 articles.

1. Intercomparison of Four Different In-Situ Techniques for Ambient Formaldehyde Measurements in Urban Air;Hak;Atmos. Chem. Phys.,2005

2. Formaldehyde and Cancer: A Critical Review;McLaughlin;Int. Arch. Occup. Environ. Health,1994

3. Cancer Effects of Formaldehyde: A Proposal for an Indoor Air Guideline Value;Nielsen;Arch. Toxicol.,2010

4. EPA, U.S. (2012). Summary of Results for the 2005 National—Scale Assessment.

5. (2023, August 17). Initial List of Hazardous Air Pollutants with Modifications | US EPA, Available online: https://www.epa.gov/haps/initial-list-hazardous-air-pollutants-modifications.

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献